High-level disinfection for endoscopy equipment

Desinfección de alto nivel para equipos de endoscopia

 

GSMED Imaging · Endoscopy
2026 Technical Guide

High-level disinfection for endoscopy equipment: protocol and products

How to ensure patient safety and regulatory compliance in endoscope reprocessing? Everything you need to know about HLD in Colombia.

HLD Protocol Glutaraldehyde · OPA Colombian Regulations Reprocessing

High-level disinfection (HLD) is the cornerstone of safety in any endoscopy unit. An improperly disinfected endoscope represents one of the most critical risks of healthcare-associated infections (HAIs), capable of transmitting multidrug-resistant bacteria, hepatitis B and C, and even Helicobacter pylori.

In this article, you will find current protocols, the most commonly used disinfectants, the difference between manual and automated processes, and the regulatory requirements set by Colombian authorities for the licensing of endoscopy services.

⚠️ Important: The Spaulding classification categorizes flexible endoscopes as semicritical devices, requiring at least high-level disinfection before each procedure. This applies to gastroscopes, colonoscopes, bronchoscopes, and any endoscope that contacts mucous membranes.

What is high-level disinfection?

HLD is a process that eliminates all pathogenic microorganisms—vegetative bacteria, mycobacteria, fungi, lipid and non-lipid viruses—except for high concentrations of bacterial spores. It is clearly distinguished from other processing levels:

Level Eliminates Endoscopy application
Sterilization Everything, including spores Reusable invasive accessories (biopsy)
HLD Bacteria, viruses, fungi, mycobacteria Flexible endoscopes (gastro, colono, bronco)
Intermediate-level disinfection Vegetative bacteria, viruses, mycobacteria External surfaces of the equipment
Low-level disinfection Vegetative bacteria, some viruses Non-critical environmental surfaces
🔬 Key point: HLD is the minimum acceptable standard for flexible endoscopes according to the ESGE-ESGENA (2018) and ASGE (2021) guidelines and the Colombian Ministry of Health. Low-level or intermediate-level disinfection is never sufficient for the body of the endoscope.

High-level disinfection protocol: step by step

The correct process always follows this sequence, without omissions. Each stage is interdependent: if the preliminary cleaning fails, the disinfectant cannot act effectively.

  1. 1. Immediate pre-cleaning (in the room) Aspirate clean water and wipe the outside of the endoscope with a damp gauze before disconnecting it. Do this within 60 seconds after removing the endoscope from the patient. This step prevents the adhesion of organic matter and biofilm formation.
  2. 2. Leak test Connect the leak tester and submerge the endoscope in water. Apply pressure and observe for bubbles for 30–60 seconds. An endoscope with a leak must not be submerged in disinfectant solution: the damage will worsen, and the endoscope must be sent for repair.
  3. 3. Manual cleaning with enzymatic detergent Submerge the endoscope in an enzymatic detergent solution (dilution according to the manufacturer). Clean the working channel with an appropriately sized brush (minimum two passes). Irrigate all channels. Rinse with clean water. This stage reduces the bioburden by > 99.9%, an essential condition for the disinfectant to be effective.
  4. 4. Immersion in high-level disinfectant Completely submerge the endoscope in the disinfectant solution (2% glutaraldehyde or 0.55% OPA). Irrigate all channels to remove air bubbles. Respect the minimum contact time established for the product at the operating temperature. Check the concentration with a test strip before each use.
  5. 5. Thorough rinsing Remove the endoscope with sterile gloves and rinse with sterile water or high-quality filtered water (water for injection, osmotic, or 0.2 µm filtered). Rinse at least three times, including all channels. Rinsing with tap water may cause recontamination with Pseudomonas or other waterborne bacteria.
  6. 6. Drying and storage Dry the outside with a sterile cloth. Blow clean air or pass 70% isopropyl alcohol through the channels to facilitate internal drying. Store vertically in a ventilated drying cabinet. Storing a damp endoscope promotes bacterial growth.
  7. 7. Traceability and recording Document: operator, date and time, endoscope (serial number), disinfectant product, batch, verified concentration, immersion time, and leak test result. This traceability is required by Resolution 2003 of 2014 for the licensing of endoscopy services in Colombia.

Most commonly used high-level disinfectants

The choice of disinfectant agent depends on compatibility with the endoscope, the time available per shift, the unit's infrastructure, and the safety profile for personnel.

Agent Concentration HLD time (20–25°C) Shelf life Key advantage
Glutaraldehyde 2% 20 min 14–28 days Low cost, widely available
OPA (orthophthalaldehyde) 0.55% 5 min (manual) · 12 min (LDE) 14 days Faster cycles, less irritation
Peracetic acid 0.2–0.35% 10–25 min (varies) Single-use or limited Excellent for automated LDEs
Glutaraldehyde + phenol According to product 20 min 28–30 days Greater stability in prolonged use

2% glutaraldehyde: the historical standard

It has been the reference agent for decades in Colombia. It requires an extraction hood because its vapors irritate the respiratory tract and skin. Personnel must wear nitrile gloves, eye protection, and a lab coat. Its advantages are its low cost and the staff's familiarity with the protocol.

OPA (orthophthalaldehyde): the modern alternative

0.55% OPA achieves HLD in 5 minutes at room temperature, compared to 20 minutes for glutaraldehyde. It has less odor and a better safety profile for personnel. Its drawback: it stains protein surfaces gray, including the skin, making glove use essential. Its cost is slightly higher than glutaraldehyde.

⚠️ Mandatory concentration verification: Before each cycle, check the minimum effective concentration with the product-specific test strips. Glutaraldehyde: minimum 1.5%. OPA: minimum 0.3%. A solution below the minimum effective concentration DOES NOT guarantee HLD, even if the immersion time is respected.

Manual vs. automated HLD: which one to choose?

Manual

Manual process

Depends on staff compliance. Requires less initial investment but greater quality control. Risk of errors in times and concentrations. Suitable for low-volume units (< 20 procedures/day).

Automated

Endoscope washer-disinfector (LDE)

Standardizes every cycle. Reduces staff exposure to chemicals. Generates electronic traceability per cycle. Higher initial investment offset by efficiency and safety. Ideal for high-volume units.

Recommended

Hybrid system

Manual cleaning (steps 1–3) + LDE for disinfection. Combines the effectiveness of manual cleaning with the standardization of the disinfection process. Preferred solution in medium-complexity hospitals in Colombia.

Automated endoscope washer-disinfectors (LDEs) must be certified according to the ISO 15883 standard and validated for the endoscopes they will process. Not all endoscopes are compatible with all LDEs: consult your supplier before acquiring the equipment.

Colombian regulations on endoscope reprocessing

Healthcare institutions (IPS) in Colombia must comply with the following regulatory frameworks for the licensing and operation of endoscopy services:

Regulatory framework Relevant aspects
Resolution 2003 of 2014 (MSPS) Conditions for licensing endoscopy services: infrastructure, equipment, personnel, and mandatory reprocessing protocols
Decree 4725 of 2005 Classification, permits, and surveillance regime for medical devices, including high-level disinfectants
INVIMA guidelines Mandatory sanitary registration for biocides (disinfectants) for hospital use. Verify the registration's validity before purchasing
ESGE-ESGENA / ASGE guidelines Adopted as technical references by the Colombian healthcare system for reprocessing protocols
NTC-ISO 15883 Standard for washer-disinfectors: water quality, cycles, validation, and maintenance
📋 Licensing checklist: To pass a licensing visit from the Health Secretariat, your endoscopy unit must have: (1) A validated reprocessing protocol, shared with the staff. (2) Traceability records for each cycle. (3) Safety data sheets for the available disinfectants. (4) Evidence of staff training. (5) Documented maintenance of the LDEs (if applicable). (6) A leak test recorded for each endoscope.

Common errors that compromise HLD

Error Consequence How to avoid it
Not checking the disinfectant concentration Ineffective HLD Colorimetric test before each cycle
Omitting the leak test Internal endoscope damage + possible contamination Mandatory leak test after each procedure
Rinsing with unfiltered tap water Recontamination with Pseudomonas Sterile or 0.2 µm filtered water
Horizontal and damp storage Bacterial growth in channels Vertical, pre-drying, ventilated cabinet
Not brushing channels between procedures Biofilm resistant to disinfectants Mechanical brushing in each cycle
Uncertified staff Process variability Documented training, assessed competency

Frequently asked questions about HLD in endoscopy

What is high-level disinfection for endoscopes?
High-level disinfection (HLD) is the process that eliminates all pathogenic microorganisms—including bacteria, viruses, fungi, and mycobacteria—except for high concentrations of bacterial spores. It is the minimum required for semicritical endoscopes according to the Spaulding classification.
How long does immersion with 2% glutaraldehyde last?
The minimum immersion time with 2% glutaraldehyde at room temperature (20–25°C) is 20 minutes for high-level disinfection. For chemical sterilization, 6 to 10 hours are required according to the product manufacturer.
What is the difference between manual and automated HLD in endoscopy?
Manual HLD depends on strict staff compliance with the protocol and is more susceptible to human error. Automated HLD using endoscope washer-disinfectors (LDEs) standardizes every cycle, reduces staff exposure to chemicals, and generates process traceability.
What Colombian regulation governs endoscope disinfection?
In Colombia, medical device disinfection is governed by Resolution 2003 of 2014 from the Ministry of Health (service licensing), INVIMA guidelines on biocides, and the international ESGE-ESGENA and ASGE guidelines are adopted as technical references.
How often should the disinfectant solution be changed?
It depends on the product: 2% glutaraldehyde can be reused for up to 14–28 days or until the colorimetric test indicates a concentration below the minimum effective level (1.5%). OPA has a shelf life of 14 days. Always check with test strips before each cycle, without exception.